FAQ & Technical support
If you have a question regarding Finapres products or services please find it in our FAQ database. If your question cannot be found please contact our customer service department. Please click here to forward your question by e-mail.  

General
Are Finapres devices compliant with MRI?
Current Finapres devices cannot be used in a MRI. The front end unit incorporates electromagnetic control valves. The valves might be affected by the strong magnetic fields in MRI devices. This will cause unreliable readings.

Can Modelflow® be used in lower ages?
This caution statement has been introduced since Modelflow® CO has been checked and published against independent other methods (thermodilution, Fick, rebreathing) in subjects over 18 years of age. It may be correct for lower ages, but there is no independent proof as yet. Modelflow® CO might be correct in paediatric patients as well, but we do not know until clinical studies have been done and published on the use of Modelflow® in lower age ranges.

What is the main differences between the Finometer® MIDI and Finometer® PRO?
The main difference is the patented RTF (Return To Flow) calibration incorporated in the Finometer® PRO. The RTF offers a higher accuracy for absolute blood pressure values. When trends are more important than absolute values the Finometer® MIDI is an outstanding choice.

What cuff sizes do you offer?
We currently offer 3 sizes cuffs; Small (Ø 45-55 mm), Medium (Ø 55-65 mm) and Large (Ø 65-75 mm). Please click here to forward a “Request for quote”.

When do I need to null (calibrate) my HCU (Height Correction Unit). How should I do that?
Hold both sensors of the HCU (transducer and reference) at the same level. If the reading is not 00 the HCU nulling procedure should be performed. Please refer to the User’s Guide for more information and about the procedure.

I can disable the “Physiocal” function. What is the influence on the measurements?
Physiocal compensates offsets due to arterial contractions/relaxations related to smooth muscle activity in the finger arteries. Switching Physiocal off has a potential effect on the setpoint (= absolute blood pressure level).

In general Physiocal can be switched off when the number of beats in between successive Physiocal interval is at its maximum step of 70, indicating that the finger physiology is stable. Usually it should not be a problem to switch Physiocal off for a period of 5 to 15 min. even when the patient is recovering from an exercise test.

Which Finapres® device is more accurate for arterial blood pressure measurement and continuous monitoring?
Our Finometer® PRO is the most accurate of our devices (most accurate device available on the market). Only the Finometer® Pro has the added feature of the patented RTF (return-to-flow) calibration which adds a level of accuracy and precision.

Commercial / Customer Service
What is the procedure and address when I would like to send my device for check up or repair?
Repairs can only be carried out in our laboratories and should always be notified to Finapres (info@finapres.com) prior to a return shipment. After receipt of a reference number the complete system, including all accessories, can be returned to Finapres Medical Systems BV, Paasheuvelweg 34a, NL-1105 BJ, Amsterdam. Please make sure the device is firmly and properly packed, if possible in the original packing, to avoid damage during transport. Please include a proforma invoice with the shipment clearly stating ‘repair return’. After receipt of the device our Customer Service will contact you.

Who should I contact for repairs?
For repair or service please contact Finapres Customer Service by email: info@finapres.com or phone: +31 20 6090974

What is your Chamber of Commerce number?
Our Chamber of Commerce number is 08200839

What are your bank account details?
Our bank account details are:
Electronic banking: ING bank account number: 65.50.37.829; swift code: INGBNL2A; IBAN: NL21INGB0655037829 or
Check: Registered mail only to Finapres Medical Systems B.V., Paasheuvelweg 34a, NL-1105 BJ, Amsterdam, The Netherlands

What is your VAT number?
Our VAT number is: NL 820931342 B01

Portapres®
Can the Portapres® be used during exercise?
The Portapres® can be used during running and cycling. However, to ensure that the measurements are correct during exercise it is very important that the hand with the finger cuffs is kept as still as possible (this could be done by using a sling).

Finometer® PRO
I need to null (calibrate) my HCU (Height Correction Unit). How should I do that?
Hold both sensors of the HCU (transducer and reference) at the same level. If the reading is not 00 the HCU nulling procedure should be performed. Please refer to the User’s Guide for more information.

Analog signals come in 3 colors. What are the colors representing?
Analog signals have the following representation:
Color brown:
(delay=0sec):extern, finger, height, armcuf, calibr, marker

Color green:
(delay=1sec):FinAP, reBAP, mFLOW

Color blue:
(delay=1sec+1beat):SYS, DIA, MAP, IBI, HR, LVET, SV, CO, TPR, Zao, Cwk, SVI, CI, PRI, PS*HR, dp/dt

The values displayed on the Finometer® screen and available on the I/O box have the same delay, but the signals you see on the Finometer® screen are the mean of 8 blood pressure measurements!
 

I am receiving the error message “check air supply”. What does this mean and what should I do?
This error indicates that there is something wrong with the air supply. It is a start up error. Please make sure that the cuff is connected and the hose is not obstructed or kinked. If the error persists please change the cuff with another cuff and try again. If the problem is solved the cuff needs to be replaced.

Finometer® MIDI
How do I connect the ECG?
Finometer® MIDI has a direct ECG input connector at the front of the device. The power voltage is supplied by the Finometer® MIDI.

I am receiving the error message “check air supply”. What does this mean and what should I do?
This error indicates that there is something wrong with the air supply. It is a start up error. Please make sure that the cuff is connected and the hose is not obstructed or kinked. If the error persists please change the cuff with another cuff and try again. If the problem is solved the cuff needs to be replaced.

I am receiving the error message “connect cuff cable”. What does this mean and what should I do?
This error indicates that the cuff is not correctly connected to your device. The following procedure can be followed to solve this issue:
1. Please disconnect and reconnect the cuff and try again 2. Change the cuff with another cuff. If the problem is solved your cuff needs to be replaced. 3. If the problem persists please try another front end or contact our customer service at info@finapres.com.

BeatScope® Software
I have problems installing BeatScope® software. What could be the cause?
The most common cause is that the computer is logged on (Windows®2000 and Windows® XT) without “Administrator Rights”. Please make sure you have “Administrator Rights” before installing BeatScope®.

I am using BeatScope® 1.0. When should I upgrade to BeatScope® 1.1a?
BeatScope® 1.0 is no longer supported. An upgrade to BeatScope® 1.1a is recommended. BeatScope® 1.1a does offer a range of additional features an improvements. Please contact our sales department for an offer.

What is Readpack?
Readpack is a software program that decompresses FM packet files. It generates 4 files: Samples file(s or .adc), Beat data file(r or .res), Header file(h or .hdr) and Text data file(a or .asc).

Ohmeda Finapres
Does Finapres Medical Systems B.V. provide support for Ohmeda Finapres systems?
The Ohmeda Finapres has been out of production and service by Ohmeda for a number of years now. We are providing replacement cuffs in the sizes small, medium and, large.

Finapres Medical Systems B.V. is able to provide the Finometer® systems to replace the Ohmeda systems. The current Finometer® systems provide a range of benefits compared with Ohmeda systems. Please click here to contact your nearest representative for more information.

Does Finapres Medical Systems B.V. supply cuffs for the Ohmeda Finapres devices?
We are providing replacement cuffs in the sizes small, medium and, large. Please click here to forward your request for quote.

Finapres Medical Systems B.V. is able to provide the Finometer® systems to replace the Ohmeda systems. The current Finometer® systems provide a range of benefits compared with Ohmeda systems.

*
*
*
*
If your inquiry is related to specific measurements or errors we generally speaking need data files (.dat or .fpf). Please upload your data file here.